Multitron - Incubation Shaker
Rectifying Faults
Error 1, Sensor
The Pt100 sensor is not returning any readings.
Error 1, Fan 1 - 4
One of the fans is blocked or defective.
Top
Bottom
In conjunction with the rotation speed parameter
(RPM symbol lights up)
Heat Error
The control of the motor has overheated because of excessively
high rotation speed or excess loading. Once the equipment has
cooled down it can be restarted.
Belt Error
The transmission belt is torn.
cntr Error
The rotation speed has exceeded the maximum allowed value
(perhaps the mainboard is defective).
Top
Bottom
In conjunction with the humidity parameter
(Humid symbol lights up)
Sensor Error
The humidity sensor is not returning any readings (humidity sen-
sor not connected or defective).
H
2
O Error
The water supply for humidification is interrupted (empty water
tank or defective hose).
Heat Error 1
The heating or temperature recording of the steam humidifier is
defective.
Heat Error 2
The heating or temperature recording of the steam humidifier is
defective.