Multitron - Incubation Shaker
Technical Data and Specifications
25 February 2019 Page 137 of 146
Target and actual humidity values
Control precision
%r H
Values that can be reached
(at 37 °C in the interior and 20 °C room
temperature)
% rH
Max. current consumption
A
Bar heater (in door)
V
W
Door seal heating
V
Specifications for the connection / water quality
Water conductivity min. 5 / max. 20 μS/cm
Dissolved solids
mg/L
Using tap water can quickly lead to limescale in the vaporiser of
the humidification unit, which would affect correct functioning.
Ultra pure water (WFI, water for injection purposes) must not be
used either as this is highly corrosive and thus damages the
equipment.
Adding detergents, disinfectants or similar chemicals to the wa-
ter for the direct steam humidification can damage the equip-
ment. Only use water without additions, in accordance to the