EasyManua.ls Logo

Infors Multitron - Page 139

Infors Multitron
149 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Multitron - Incubation Shaker
Technical Data and Specifications
25 February 2019 Page 137 of 146
Target and actual humidity values
Description
Value
Unit
Setting range
20.0 to 85.0
% rH
Increment
1.0
% rH
Control precision
± 3.0
%r H
Values that can be reached
(at 37 °C in the interior and 20 °C room
temperature)
85.0
% rH
Electrical data
Description
Value
Unit
Voltage
230
V
Max. power consumption
125
W
Max. current consumption
0,4
A
Bar heater (in door)
24
V
15
W
Door seal heating
24
V
17
W
Floor heating
24
V
17
W
Specifications for the connection / water quality
Description
Value
Unit
Connection (hose Ø)
6 to 7
mm
Input pressure
0 to 0.3
bar
Water hardness
< 0.01, CaCO
3
equivalent
mmol/L
Water conductivity min. 5 / max. 20 μS/cm
Dissolved solids
< 10
mg/L
ATTENTION
Using tap water can quickly lead to limescale in the vaporiser of
the humidification unit, which would affect correct functioning.
Ultra pure water (WFI, water for injection purposes) must not be
used either as this is highly corrosive and thus damages the
equipment.
Adding detergents, disinfectants or similar chemicals to the wa-
ter for the direct steam humidification can damage the equip-
ment. Only use water without additions, in accordance to the
specification.

Table of Contents