Multitron - Incubation Shaker
Options
25 February 2019 Page 35 of 146
The equipment can be equipped with an internal cooling unit. The
cooling unit can either be placed in a bracket on the incubation
shaker or in the high base (900 W version only).
The cooling unit is available in two different versions:
Version with 900 W cooling capacity (one compressor)
Version with 1800 W cooling capacity (two compressors)
The version with 1800 W cooling capacity is used when very low
temperatures have to be reached. As two independent compres-
sors are installed in this version, it is also ensured that cooling can
continue if one compressor fails (redundancy).
The cooling unit is an autonomous system and has its own power
supply and mains switch. This is located on the right-hand side of
the bracket or base. The cooling unit is a 2-circuit system with a
coolant tank. Prior to commissioning, it must be filled with coolant
by the manufacturer’s qualified expert personnel or persons au-
thorised by the manufacturer.
The heat exchanger is located behind the rear wall in the incuba-
tion chamber.