Multitron - Incubation Shaker
Installation
Page 68 of 146 25 February 2019
Placing the equipment under a table
When placing the equipment under a table, observe the following:
You must leave a ventilation gap of at least 100 mm between
the equipment and rear wall of the table to ensure heat pro-
duced by the equipment can escape.
The minimum distance between the equipment and the un-
derside of the table top is 100 mm.
When operating the equipment with installed cooling, heat
can accumulate under a closed laboratory bench. This can
reduce the temperature range of the equipment and result in
increased power consumption. Hence, we recommend only
installing the equipment under a laboratory bench with air
vents.
Failure to comply with the required minimum distances can re-
sult in components of the equipment overheating and the equip-
ment being damaged. Therefore, please note:
– Never cover the air vents on the right side or those on the
rear of the equipment.
– Never position the equipment directly on a wall.