SYSTEM OVERVIEW
Manual Part Number
14
Cradle Type: 3.0 (Prostate)
Software version: 8.1
System Specifications:
Transducer: 990 transmit elements flat phase array
Transducer aperture (diameter): 34x22mm
Focal distance: 15-60mm (achieved by electronic steering)
Frequencies: 2.3±0.25Mhz
Maximal Energy per
macro-spot: 30KJ
Maximal energy per sub-spot: 2.2KJ
Energy specifications:
Energy density in target: 400~800J/cm
2
Ultrasound output: ~45W
Max output pulse width: 512sec
Effective radiating area (ERA): 748mm
2
Focal region size: Variable
Focal control: Electronic
Tissue destruction mechanism: Thermal coagulative necrosis
Imaging during treatment: PRF MR Images at 3-6seconds interval
Temperature measurement: Proton Resonance Frequency (PRF) MR thermometry
Reflection monitoring: None
Cavitation monitoring: Real time display
Treatment outcome monitoring method: Using real-time thermometry the operator gets real time
feedback on treatment outcome, as temperature is correlated to tissue viability.
Independent assessment using standard MR imaging sequences is done during and after treatment,
to assess lesion size and location.
System Electrical Specifications:
Rated power voltages/currents: 208V (30A) / 380-400V (15A) / 480V (13A)