WATER SYSTEM MAINTENANCE
Manual Part Number
158
10.
High flow chiller
The coolant flow rate in the is
above threshold.
Check for coolant leaks around the table
and along the hoses up to the chiller (in
equipment room).
Check visually for bubbles in the coolant,
the error might reoccur as long as there
are bubbles in the coolant.
11.
Low flow chiller
The coolant flow rate in the is
below threshold.
Check the water hose connection to the
table.
Check visually for bubbles in the coolant,
the error might reoccur as long as there
are bubbles in the coolant.
12.
High/Low SU Temp
Temperature in the
Switching Unit (on patient
table) is not within limits.
Click Reset on the remote control.
If the problem persists, contact technical
support.