10-10 SCPI Command Reference Models 2510 and 2510-AT User’s Manual
:SOURce[1]
:TEMPerature
:PROTection Set temperature protection. ✓
[:HIGH] Upper temperature limit. ✓
[:LEVel] <n> Set upper temperature limit.
<n> = -50 to +250°C, 223.15 to 523.15K,
-58 to 482°F | MIN | MAX | DEF
50.0°C,
323.15K,
122°F
✓
[:LEVel]? Query upper protection temperature limit. ✓
:TRIPped? Query if upper limit is exceeded. (1 = yes;
0 = no).
✓
:LOW Lower temperature limit. 0.0 ✓
[:LEVel] <n> Set lower temperature limit.
<n> = -50 to +250°C, 223.15 to 523.15K,
-58 to 482°F | MIN | MAX | DEF
0°C,
273.15K,
32°F
✓
[:LEVel]? Query lower protection temperature limit. ✓
:TRIPped? Query if lower limit exceeded. (1 = yes;
0 = no).
✓
:STATe <b> Enable/disable temperature protection.
(ON | OFF)
ON
:STATe? Query temperature protection state.
[:SPOint] <n> Program temperature setpoint.
<n> = -50 to +225°C, 223.15 to 498.15K,
-58 to 437°F | MIN | MAX | DEF
25°C,
298.15K,
77°F
✓
[:SPOint]? Query temperature setpoint. ✓
Table 10-5 (continued)
:SOURce[1] subsystem commands
Command Description Default parameter SCPI