EasyManua.ls Logo

Keithley 2657A

Keithley 2657A
805 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
A
ppendix B: Calibration Model 2657A High Power System SourceMeter® Instrument Reference Manual
B-18 2657A-901-01 Rev. B/December 2012
Model 2657A calibration adjustment steps
Function
1
Adjustment
steps
2
Adjustment
points
3
Sense mode
4
Voltage source and measure
See Step 2. Voltage calibration
adjustment (on page B-20)
200 V ±1e-30, ±180 V
smua.SENSE_LOCAL
200 V ±1e-30, ±180 V
smua.SENSE_REMOTE
200 V ±1e-30, ±180 V
smua.SENSE_CALA
500 V ±1e-30, ±450 V
smua.SENSE_LOCAL
500 V ±1e-30, ±450 V
smua.SENSE_REMOTE
1500 V ±1e-30, ±1350 V
smua.SENSE_LOCAL
1500 V ±1e-30, ±1350 V
smua.SENSE_REMOTE
1500 V ±1e-30, ±1350 V
smua.SENSE_CALA
3000 V ±1e-30, ±2700 V
smua.SENSE_LOCAL
3000 V ±1e-30, ±2700 V
smua.SENSE_REMOTE
3000 V ±1e-30, ±2700 V
smua.SENSE_CALA
Current source and measure
See Step 3. Current calibration
adjustment (on page B-25)
1 nA
±1e-30, ±0.9 nA
smua.SENSE_LOCAL
10 nA ±1e-30, ±9 nA
smua.SENSE_LOCAL
100 nA ±1e-30, ±90 nA
smua.SENSE_LOCAL
1 μA ±1e-30, ±0.9 μA
smua.SENSE_LOCAL
10 μA ±1e-30, ±9 μA
smua.SENSE_LOCAL
100 μA ±1e-30, ±90 μA
smua.SENSE_LOCAL
1 mA ±1e-30, ±0.9 mA
smua.SENSE_LOCAL
1 mA ±1e-30, ±0.9 mA
smua.SENSE_CALA
2 mA ±1e-30, ±1.8 mA
smua.SENSE_LOCAL
20 mA ±1e-30, ±18 mA
smua.SENSE_LOCAL
120 mA ±1e-30, ±108 mA
smua.SENSE_LOCAL
Internal calibration
adjustments
See Step 4. Internal calibration adjustments (on page B-28)
Contact check See Step 5. Contact check calibration adjustment (on page B-29)
1
Adjust only the source for the SENSE_CALA sense steps.
2
Steps must be performed in the order shown.
3
Do not use actual 0 values for zero calibration adjustment points. Use very small values such as ±1e-30.
Calibration adjustment polarities must also be set as shown in the procedures.
4
Output must be off before changing to the CALA sense mode.

Table of Contents

Related product manuals