Keysight E5260/E5270 User’s Guide, Edition 6 4-29
Front Panel Reference
Setup Menus
DIAG Displays the sub menu that provides the following functions.
EXECUTE
Displays the following message.
DIAG: Item
where Item is one of the following:
GPIB GPIB diagnostics
TRIGGER (BNC) Trigger In/Out diagnostics
KEY Front panel key diagnostics
HV-LED High voltage LED diagnostics
DGTL-I/O Digital I/O diagnostics
BEEPER Beeper diagnostics
Press the arrow key and the
Enter
key to select the diagnostics item, and refer to
Table 4-5 to perform the diagnostics; or press the
Exit
key to return to the
previous menu without performing diagnostics.
After the diagnostics, the following message is displayed:
DIAG: Pass or Fail or Aborted
Press the
Exit
key to return to the previous menu.
RESULT
Displays the following message.
DIAG: Item Pass or Fail Error No. or Not performed yet
where Item is the diagnostics item and Error No. is the error number.
Press the arrow key to display the result of another item.
Press the
Exit
key to return to the previous menu.

Table of Contents

Other manuals for Keysight Technologies E5260 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies E5260 Series and is the answer not in the manual?

Keysight Technologies E5260 Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelE5260 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals