User Permission of Viewing Errors Fault Description Corresponding to the
Permission
User and Administrator Permission
GPU temperature alarm. Please save the
current data quickly. Shut down in xx seconds.
Service Engineer Permission GPU temperature alarm. Please save the
current data quickly. Shut down in xx seconds.
Log Record
"Date" "Time" "Errocode": System Monitor:Temperature alert! [XXX], Current temperature: [VVV]
C, Current Fan Speed:[FFF]rpm
Involved FRU
GPU module assembly
Severity
Suspension
Handling Level
Contact the device provider for onsite handling.
Error Mechanism
The system detects that the GPU temperature exceeds the threshold, triggering automatic
shutdown.
Possible Cause
1. The ambient temperature for the device is too high.
2. There are sundries around the air inlet and outlet of the device, causing poor ventilation and
dissipation.
3. The dust-proof filter inside the device is blocked, which prevents heat from being discharged.
4. A corresponding component of the device is faulty.
Resolve Path
Confirm the ambient temperature and ventilation and dissipation of the device and clean the dust-
proof net. If the error persists, replace the corresponding component inside the machine.
Solution
Diagnostic Ultrasound System
Service Manual
8 Troubleshooting