340
1. If the ambient temperature is too high, move the machine to an area with a lower ambient
temperature (such as an air-conditioned room).
2. Remove the sundries from the air inlet and outlet of the device to ensure good ventilation and
dissipation.
3. Remove the dust-proof filter inside the device and perform cleaning and maintenance.
4. If the fault persists after the preceding measures are taken, replace the GPU module. For
details, see 7.12.2
Disassembly and Assembly.
88..22..33..88 T209: SSD temperature alarm. Please save the current data quickly.
Shut down in xx seconds.
Failure Code
T209
Fault Description
User Permission of Viewing Errors Fault Description Corresponding to the
Permission
User and Administrator Permission
SSD temperature alarm. Please save the
current data quickly. Shut down in xx seconds.
Service Engineer Permission SSD temperature alarm. Please save the
current data quickly. Shut down in xx seconds.
Log Record
"Date" "Time" "Errocode": System Monitor:Temperature alert! [XXX], Current temperature: [VVV]
C, Current Fan Speed:[FFF]rpm
Involved FRU
SSD assembly
Severity
Suspension
Handling Level
Contact the device provider for onsite handling.
Error Mechanism