8.3.3 System Startup Errors
88..33..33..11 When the on/off button is pressed and the main screen displays
"Diagnostic Ultrasound System", but the system cannot continue
Fault Description
When the on/off button is pressed and the main screen displays "Diagnostic Ultrasound System",
but the system cannot continue
Involved FRU
PC module and SSD
Severity
Suspension
Handling Level
Contact the device provider for onsite handling.
Possible Cause
The system powers up normally and enters BIOS self-test, but fails to continue.
Resolve Path
1. Disconnect all peripheral devices and check whether the fault persists. If so, go to Step 2.
2. Disconnect SSD and check whether the system is stuck on the Diagnostic Ultrasound System
screen. If so, the PC module is faulty. If not, the SSD card is faulty.
Solution
Replace the PC module or SSD card. For details, refer to 7.11.2 Disassembly and Assembly or
7.15.2
Disassembly and Assembly.
88..33..33..22 The system displays "Diagnostic Ultrasound System" and then the
screen goes black
Fault Description
The system displays "Diagnostic Ultrasound System" and then the screen goes black
Involved FRU
Diagnostic Ultrasound System
Service Manual
8 Troubleshooting

Table of Contents

Other manuals for Mindray TEX20

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray TEX20 and is the answer not in the manual?

Mindray TEX20 Specifications

General IconGeneral
BrandMindray
ModelTEX20
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals