SSD
Severity
Suspension
Handling Level
Contact the device provider for onsite handling.
Possible Cause
Failed to load the operating system. The SSD card or the operating system root directory is
damaged.
Resolve Path
Restore the operating system and Doppler software. If the fault persists, replace the SSD card.
Solution
Restore the operating system and Doppler software. If the fault persists, replace the SSD card.
For details, refer to 7.15.2
Disassembly and Assembly.
88..33..33..33 The system displays the Doppler startup screen, but no further step
can be performed.
Fault Description
The system displays the Doppler startup screen, but no further step can be performed.
Involved FRU
SSD
Severity
Suspension
Handling Level
Contact the device provider for onsite handling.
Possible Cause
Failed to load the Doppler software. The SSD card or the operating system in it is damaged.
Diagnostic Ultrasound System
Service Manual
8 Troubleshooting

Table of Contents

Other manuals for Mindray TEX20

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray TEX20 and is the answer not in the manual?

Mindray TEX20 Specifications

General IconGeneral
BrandMindray
ModelTEX20
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals