Page #392 background imageLoading...
Page #392 background image
Revision:1.0(2023-01-12)
393
2. Use a new probe to conduct a cross test. If the fault does not appear on the new probe, replace
the probe.
3. If the above factors are eliminated, replace the main board. For details, refer to 7.13.2
Disassembly and Assembly.
8.3.5 Probe Recognition Errors
88..33..55..11 A Single Probe Cannot Be Recognized by Any Probe Port
Fault Description
A Single Probe Cannot Be Recognized by Any Probe Port
Involved FRU
Probes
Severity
Suspension
Handling Level
Contact the device provider for onsite handling.
Possible Cause
1. The system does not have the POD data for this probe model.
2. The probe is faulty.
Resolve Path
1. Press Setup on the touch screen to enter the Setup interface. Choose
System>Image>RESET CONFIG, and check whether the probe model can be found in the
Probe drop-down list. If not, the system does not have the POD data for this probe model.
2. If the POD data exists, the probe may be faulty.
Solution
1. If the POD data is missing, restore the Doppler software to the latest version.
2. If the POD data exists, replace the probe.
88..33..55..22 No Probes Can Be Recognized
Fault Description
Diagnostic Ultrasound System
Service Manual
8 Troubleshooting

Table of Contents

Other manuals for Mindray TEX20

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray TEX20 and is the answer not in the manual?

Mindray TEX20 Specifications

General IconGeneral
BrandMindray
ModelTEX20
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals