Revision:1.0(2023-01-12)
394
No Probes Can Be Recognized
Involved FRU
Probe board and main board
Severity
Suspension
Handling Level
Contact the device provider for onsite handling.
Possible Cause
1. The probe board is faulty (low possibility).
2. The main board is faulty.
3. The system does not have the POD data for this probe model (low possibility).
Resolve Path
1. Clean the probe socket port, and connect and test the probe again. If the fault persists, replace
the main board first.
2. Press Setup on the touch screen to enter the Setup interface. Choose
System>Image>RESET CONFIG, and check whether the probe model can be found in the
Probe drop-down list. If not, the system does not have the POD data for this probe model.
Solution
1. Clean the probe adapter port, and connect and test the probe again. If the fault persists,
replace the main board first. For details, refer to 7.13.2
Disassembly and Assembly.
2. If the fault persists after the main board is replaced, replace the probe board. For details, refer
to 7.9.2
Disassembly and Assembly.
8.3.6 Display and Related Errors
88..33..66..11 The device startup is normal, a black screen occurs on the main
display, and the display indicator is green.
Fault Description
The device startup is normal, a black screen occurs on the main display, and the display indicator
is green.
Diagnostic Ultrasound System
Service Manual
8 Troubleshooting