A-9
ST Low
-2.0 mV to (high limit - 0.2 mV)
-2.0 mV to 2.0 mV
A.7.2 Resp
Technique Trans-thoracic impedance
Lead Options are lead I and II. The default is lead II.
Respiration excitation
waveform
<300 μA RMS, 64 kHz (±10%)
Baseline impedance range 200 to 2500 (using an ECG cable with 1k resistance)
Bandwidth 0.2 to 2.5 Hz (-3 dB)
Sweep speed
3 mm/s, 6.25 mm/s, 12.5 mm/s, 25 mm/s, or 50 mm/s
Accuracy: ±5%
Respiration Rate
Measurement range
Adult:
Pediatric, neonate:
0 to 120 rpm
0 to 150 rpm
Resolution 1 rpm
Accuracy
7 to 150 rpm:
0 to 6 rpm:
±2 rpm or ±2%, whichever is greater
Not specified.
Apnea alarm time 10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s
Alarm limit Range (rpm) Step (rpm)
RR High
Adult, pediatric:
Neonate:
(low limit + 2) to 100
(low limit + 2) to 150
1
RR Low 0 to (high limit – 2)