EasyManua.ls Logo

Mindray uMEC15 - Page 255

Mindray uMEC15
288 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
C-7
C.1.6 NIBP
Item Name
O.M
General User Defaults
C M
Alarm * * On
Alm Lev * * Med
Interval * * 15 min
NIBP End Tone * * Off
Cuff Press.
˄mmHg˅
Adu
* *
80
Ped 60
Neo 40
Initial
Pressure
˄mmHg˅
Adu
* *
160
Ped 140
Neo 90
Alarm Limits
NIBP-S High
˄mmHg˅
Adu
* *
160
Ped 120
Neo 90
NIBP-S Low
˄mmHg˅
Adu
* *
90
Ped 70
Neo 40
NIBP-M High
˄mmHg˅
Adu
* *
110
Ped 90
Neo 70
NIBP-M Low
˄mmHg˅
Adu
* *
60
Ped 50
Neo 25
NIBP-D High
˄mmHg˅
Adu
* *
90
Ped 70
Neo 60
NIBP-D Low
˄mmHg˅
Adu
* *
50
Ped 40
Neo 20

Table of Contents

Related product manuals