C-8
C.1.7 IBP (for uMEC12/uMEC15)
Item Name
O.M
General User Defaults
C M
IBP 1 Label * * Art
IBP 2 Label * * CVP
Alarm * * On
Alm Lev * * Med
P1 Measure * * All
P2 Measure * * All
P3 Measure * * Mean
P4 Measure * * Mean
Sensitivity * * Med
Sweep * * 25 mm/s
Filter * 12.5 Hz
Gridlines * * Off
IBP Label Order
Setup
* *
Art, pArt, CVP, pCVP, ICP, PA, Ao, UAP, FAP,
BAP, LV, LAP, RAP, UVP, P1, P2, P3, P4
Art, Ao, UAP, BAP, FAP, LV, P1-P2 Arterial Pressure Alarm Limits
IBP-S High
˄mmHg˅
Adu
* *
160
Ped 120
Neo 90
IBP-S Low
˄mmHg˅
Adu
* *
90
Ped 70
Neo 55
IBP-M High
˄mmHg˅
Adu
* *
110
Ped 90
Neo 70
IBP-M Low
˄mmHg˅
Adu
* *
70
Ped 50
Neo 35
IBP-D High
˄mmHg˅
Adu
* *
90
Ped 70
Neo 60
IBP-D Low
˄mmHg˅
Adu
* *
50
Ped 40

Table of Contents