52
7−4−3.
Target of Zero tracking
Zero tracking of this instrument can select the target from either “Gross weight” or “Net
weight”.
This selection is made in Function mode(related Function/FUNC−62).
As for default, “gross weight” has set.
7−4−4. Cancellation for compensation by zero tracking.
Cancellation for data cancelled by zero tracking can be executed with Function/FUNC−98.
When changing the target for zero tracking is required, be sure to
cancel the compensated portion once that has compensated by zero
tracking with FUNC−98.