25
5−2−2.
Fine adjustment mode.
During tare cancellation(A/Z) and zero set is executed, and during
zero tracking is effective, fine adjustment mode can’t be entered.
(Displays “ERR−5”)
When HOLD signal is input with the display or with the condition of
target of hold(FUNC−50) for option, fine adjustment mode can’t be
entered. When fine adjustment is made, be sure to cancel the HOLD
input.
Procedures
With the key pressed, press the key
together for 2 seconds.
Fine adjustment mode can be entered, and
present load value on the load display will flash
on and off.
The S1 display will show the counter value for
coarse adjustment.
S3 display will show “ZERO”
S2 will show up to 1/10 of present load value.
S4 will show the value (A/D value) operated
internally.
1
When fine adjustment on load is required,
complete the calibration on CCAL mode
beforehand.
In fine adjustment mode, cancellation of
difference can be made, which has occurred
due to the difference between calibration
data and fundamental data of the
instrument when actual load is applied.
In case of interruption in fine adjustment,
Measurement mode can be re−entered by
pressing the key.
In this case, up to the last data will be
effective.
2
Set the instrument with initial load application.