150
10−2. Error display
Error Code Contents of Error Remedy
ERR−0 Execute zero set exceeding ±10 % of
maximum display value.
Apply zero set after making within ±10 %
of maximum display value.
ERR−1 Mistake in setting Set correctly.
ERR−2 Mistake in setting during calibration. Set correctly.
ERR−3 A/D error Turn off the power and supply the power
again. If Error is still displayed, contact
with Minebea.
ERR−5 Displays when tare, zero set and zero
tracking are ON during fine adjustment
on load.
Set the Tare cancellation(A/Z OFF), ZERO
set clear(FUNC−98), and zero tracking
OFF. Set the FUNC−07 to “000000”.
TER−L Displays when Tare is less than
−2.4 mV during calibration.
TER−H Displays when Tare is greater than
2.4 mV.
ERR−E EEPROM error
ERR−R EEPROM error
ERR−B Battery is run out. Contents of memory
has destroyed.
Press the key.
Measurement mode can be entered.
(This time, A/Z data becomes initial value.)
If the Error display shows every time when
power is ON, contact with Minebea.
HCAL Initial calibration has not completed or
initial data has broken.
Contact with Minebea.