EasyManua.ls Logo

Mitutoyo SJ-301 - Page 176

Mitutoyo SJ-301
346 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5 - 78
No. 99MBB091A
12
<Tilt Compensation Arbitrary Setting
Input screen>
13
<Measurement Condition Setup screen>
TiltCompensation
P2
4 5 6
7BS
AC
8 9
1 2
0
3
ESC
mm
0.00
.
(0.00 99.99mm)
Condition menu
RE-CAL.
0.00-09.00m
90.0-99.00mm
ON
Page3/3
DRIVE
OFF
11
<Tilt Compensation P4 Input screen>
Tilt compensation
mm
mm
mm
mm
P3
P4
0.00
P2 9.00
9.00
P1 0.00
ESC
P1 P2 P3 P4
P1 P2 P3 P4
P1 P2 P3 P4
Enter the 4th point in the tilt compensation range in the Tilt
Compensation P4 Input field, and then touch the [Enter] key.
The P4 enterable range is as shown below.
0.00 to 99.99mm (in the units of 0.01mm) as tilt compensa-
tion setting
NOTE The P4 value must not be less than the P3
value.
TIP If the P2 value is larger than the evaluation
length, the end of a measured profile be-
comes the end point of tilt compensation.
For information about how to enter a numeric
value, refer to “2.2 Outline of the Touch
Panel Inputting numeric values”.
The Tilt Compensation Input screen is restored.
Touch the [Enter] button.
The tilt compensation range is loaded, and the Measure-
ment Condition Setup screen is restored.
Touch the [RE-CAL] button.
Recalculation is performed and the Home screen is
restored.

Table of Contents

Related product manuals