MOOR INSTRUMENTS moorLDI2 RESEARCH USER MANUAL
121
External Surfaces and Cables
The mobile stand, scan-head enclosure (but
NOT
the Optic window) and cabling can be disinfected
using an approved medical wipe (e.g. Medi-Wipe
) containing:
Isopropanol (rubbing alcohol) or Ethanol alcohol >=70% w/w.
And an optional antiseptic agent such as Chlorhexidine Digluconate.
Guide for using medical wipes:
Turn the system off and disconnect the mains cable before cleaning.
Ensure that the medical wipes are not overly wet when cleaning the enclosure.
Allow the system to dry fully before connecting the mains cable and switching on.
Optic Window
If dust settles on the optic window this can be removed using low pressure compressed air, e.g.
using a commercially available compressed air canister.
Do not attempt
to wipe the surface of
the optic window because it is easily scratched.
Cleaning an optical surface, and in particular a glass surface or front silvered mirror, other than
with compressed air should only be done by someone with appropriate experience and skills.
20.7 CALIBRATION AND STABILITY
Calibration and stability testing is described in Section 19 and should be done at least on an annual
basis and following impacts which might occur during transport.
We recommend that the calibration software is used weekly to assess if calibration is required.