MOOR INSTRUMENTS moorLDI2 RESEARCH USER MANUAL
10
1.6 ENVIRONMENTAL CONDITIONS
The scanner should not be used in conditions of high humidity (>80%) as this could cause misting
(condensation) on the optical surfaces e.g. mirrors and aperture window. Also it is a potential
electrical hazard.
The room temperature in which the scanner is operated must be in the range 15
o
C to 30
o
C. The
instrument storage temperature range is 0-45
o
C.
The operator should ensure the scanner is securely supported and free from vibrations due to external
influences.
1.7 WARNINGS / CAUTION DO'S AND DON'TS
Do turn off mains power before disconnecting or connecting leads - possibility of damage to
electrical components and corrupting the software.
Do turn on the computer before turning on the scanner control unit - possible loss of control of the
scanner requiring power to be switched off and then re-powering in correct sequence.
Do follow the instructions for the safe operation of the laser - see Section 1.2 of this manual.
Do not touch the transparent area of the optical window.
Do not attempt to clean the optical window unless you have been instructed how to do so and have
been authorised to do so - the surface is easily scratched. If scratched it will have to be replaced.
Do consult the responsible person for the moorLDI2 in your establishment or a Moor Instruments
representative if you need advice.
WARNING: To avoid the risk of electric shock, this equipment must only be connected to a supply
mains with protective earth.
WARNING: No modification of this equipment is allowed.
WARNING: The front panel of the scan head contains glass which is susceptible to shattering on