MOOR INSTRUMENTS moorLDI2 RESEARCH USER MANUAL
134
APPENDIX 1. PREPARING YOUR moorLDI2 IMAGER FOR THE NEW
SOFTWARE
The following procedures have been performed at manufacture and installation for all
moorLDI2
imagers from August 2002. The proceedures should be applied to earlier
moorLDI2
imagers.
Overview
Before you can run the latest release software, it is necessary to:
Use the Service Software (mldise29u) to record the serial number of your system in LDI memory.
Use the Calibration Software V1.1 to perform a calibration check and recalibrate if required.
Use the Service Software (mldise29u) to calculate the individual distance scaler for your system.
Note: only one software program should be running at any one time:
Close the Service software prior to opening the Calibration software.
Close the Calibration software prior to opening the Service software.
When these tasks have been completed, you should use the installation disc supplied with this package to load
the new software onto your PC.
Full information is to be found on the following pages.