3.7 BEFORE FIRST USE
Once you have installed the software and set up the moorLDI2 system, you must carry out the
following:
•
Visual inspection of the whole system, looking for signs of damage or missing components.
•
Turn the system on and carry out a system calibration check following the procedure detailed
in section “19 Calibration and Stability”, of this user manual
•
Check that the distance measurement is accurate to ± 5cm at 80cm scan distance
If you suspect your system may have been damaged, or if your system fails the calibration or distance
checks, please contact your distributor or Moor Instruments directly, quoting the instrument serial
number