21. ELECTROMAGNETIC COMPATABILITY
The moorLDI2 laser Doppler imager (SN5500 or above) complies with the requirements of IEC
60601-1-2:2007 Electromagnetic Compatibility of Medical Electrical Equipment. Medical electrical
equipment needs special precautions regarding electromagnetic compatibility and needs to be
installed and put into service in accordance with the following information.
Mobile radio frequency communications equipment can affect medical electrical equipment.
The moorLDI2 should not be used adjacent to or stacked with other equipment. If adjacent or
stacked use is necessary, the moorLDI2 should be observed to verify normal operation in the
configuration in which it will be used.
21.1 CABLING
The moorLDI2 is compliant with IEC 60601-1-2:2007 only when used with the following range of
connecting leads. Modification of cables or use of cables other than those supplied may result in
increased emissions or decreased immunity of the equipment.
PORT 1 LEAD
25 way lead connecting the control box to scanner head (2 or 4 meters).
PORT 2 LEAD
8 way lead connecting the control box to scanner head (2 or 4 meters).
VIDEO LEAD
coaxial video lead connecting the control box to scanner head (2 or 4 meters).
RS232 LEAD
RS232 lead connecting the control box to PC (less than 3 meters).
FIREWIRE LEAD
Firewire 6-4 or 6-6 way lead connecting the control box to PC (less than 3
meters).
21.2 GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC
EMISSIONS
Guidance and manufacturer’s declaration – electromagnetic emissions
The moorLDI2 is intended for use in the electromagnetic environment specified below. The user
of the moorLDI2 laser Doppler imager should assure that it is used in such an environment.
Emissions Test Compliance Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The moorLDI2 uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B
The moorLDI2 is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
(Power input <
75W)
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Not applicable
(Power input <
75W)