100 Communication Command Reference
8.3.3.54 PM:MAX:Power?
Description Maximum power query
Syntax PM:MAX:Power?
Remarks The PM:MAX: Power? returns current range’s maximum
readable power.
Response Type Description
Power float Power in Watts
8.3.3.55 PM:MIN:Lambda?
Description Minimum wavelength query
Syntax PM:MIN:Lambda?
Remarks The PM:MIN:Lambda? query returns the shortest
calibrated wavelength in nanometers. If no detector is
preset,min lambda for the last read detector is returned.
Response Type Description
Wavelength int Wavelength in nanometers (nm)
Related Commands: PM:MAX:Lambda?
8.3.3.56 PM:MIN:Power?
Description Minimum power query
Syntax PM:MIN:Power?
Remarks The PM:MIN: Power? returns current range’s minimum
readable power.
Response Type Description
Power float Power in Watts
Related Commands: PM:MAX:Power?
8.3.3.57 PM:MEAS:TIMEOUT
Description Measurement timeout set command
Syntax PM:MEAS:TIMEOUT value
Remarks This command sets the measurement timeout period. This
value is used for making the following measurements:
1. Peak-to-peak measurements: These measurements are
updated once every timeout period when the power meter is
in Peak-Peak Continuous mode.

Table of Contents

Other manuals for Newport 1936-R series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Newport 1936-R series and is the answer not in the manual?

Newport 1936-R series Specifications

General IconGeneral
BrandNewport
Model1936-R series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals