58 System Operation
4.9 Measurement Correction Settings
The 1936/2936 Series power meters provide users the capability to correct
actual measurements taken by it through a “Correction Settings” screen.
This screen can be accessed by pressing the “Correction” soft-key in
“Measurement Settings” screen (Figure 51). The “Measurement Settings”
screen, as described earlier, can be accessed by pressing the Setup key
from main measurement screen.
The Correction Settings screen
(Figure 52) allows users to enter
two (2) multiplier values, labeled
“Multiplier 1” (M1) and
“Multiplier 2” (M2), besides an
“Offset” value. The corrected
measurement is arrived at using the
formula shown below:
Corrected measurement = [(Actual
measurement * M1) Offset] *
M2.
Note that these settings are channel
specific, and apply to the primary
channel viewed on the main
measurement screen.
The default value for the two
multipliers is 1.0, and the offset is
0.0. If any one of these three
parameters are changed from their
default values, the main
measurement screen shows the
corrected measurement. A new field
called “Detector:” appears at the
bottom of the main screen that
displays
Figure 51 Measurement Settings
Figure 52 Correction Settings

Table of Contents

Other manuals for Newport 1936-R series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Newport 1936-R series and is the answer not in the manual?

Newport 1936-R series Specifications

General IconGeneral
BrandNewport
Model1936-R series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals