Communication Command Reference 83 83
Syntax BRIGHT brightness
Remarks The BRIGHT command controls the brightness of the
instrument display.
Argument Type Description
Brightness int Brightness, in levels from 0 to 7
Related Commands: DISP:BRIGHT?
8.3.2.2 DISP:BRIGHT?
Description Display brightness query
Syntax DISP:BRIGHT?
Remarks The BRIGHT? query returns the display brightness setting.
Response Type Description
Brightness int Display brightness, in levels from 0 to 7
Related Commands: DISP:BRIGHT
8.3.3 Power Meter Commands
8.3.3.1 *IDN?
Description Identification Query
Syntax *IDN?
Parameters None
Function
Remarks This query will cause the power meter to return an identification string.
NEWPORT XXXX vYYY mm/dd/yy SNZZZZ
Examples:
NEWPORT 1936-R v1.0.0 12/12/05 SN0001
NEWPORT 2936-R v1.0.0 12/12/05 SN0001
8.3.3.2 *RCL
Description Recall Configuration Settings
Syntax *RCL bin
Remarks The *RCL command restores the power meter to the setup
state saved in its non-volatile flash memory.