EasyManuals Logo

Newport 1936-R series User Manual

Newport 1936-R series
150 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #110 background imageLoading...
Page #110 background image
110 Communication Command Reference
Function
This query returns a value showing whether the external trigger input is
enabled or disabled.
Returns <state>
<state>is of type <number>that represents the integer 0,if the external trigger
input is disabled. If <state> is 1, 2, or 3, the external trigger is enabled on
channel A, channel B, or both respectively.
Related Commands: PM:TRIG:EXTernal,
8.3.3.87 PM:TRIG:EDGE
Description External Trigger Edge Select
Syntax PM:TRIG:EDGE <edge>
Parameters
The parameter <edge>is either 0 or 1.If <edge>is 0 then the external trigger is
defined as a falling edge. If <edge>is 1 then the external trigger is defined as
a rising edge.
Function
This command defines whether the external trigger input on the back panel is
falling edge or rising edge active.
Related Commands: PM:TRIG: EXTernal, PM:TRIG:EXTernal?,
PM:TRIG:EDGE?
8.3.3.88 PM:TRIG:EDGE?
Description External Trigger Edge Query
Syntax PM:TRIG:EDGE?
Parameters None
Function
This query returns a value showing which external trigger edge is the active
edge.
Returns <edge>
<edge>is of type <number>and represents 0,if the external trigger is defined
as a falling edge, or 1,if the external trigger is defined as a rising edge.
Related Commands: EXT PM:TRIG: EXTernal, PM:TRIG:EXTernal?,
PM:TRIG:EDGE
8.3.3.89 PM:TRIG:HOLDoff
Description External Trigger Holdoff Time Command

Table of Contents

Other manuals for Newport 1936-R series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Newport 1936-R series and is the answer not in the manual?

Newport 1936-R series Specifications

General IconGeneral
BrandNewport
Model1936-R series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals