Communication Command Reference 113 113
Syntax PM:TRIG:STATE <option>
Parameters
The parameter <option> is of type <number> and is an integer 0 or 1.
Argument Value Description
option
0
Trigger is armed. The system waits for a
trigger start event to occur in order to make a
measurement.
1
System is triggered and at least one
measurement occurred.
Related Commands: PM:TRIG: STATE?
8.3.3.96 PM:TRIG:STATE?
Description Trigger STATE Query
Syntax PM:TRIG:STATE?
Parameters None
Function
This query returns the TRIGGER STATE condition.
Related Commands: PM:TRIG: STATE
8.3.3.97 PM:TRIG:VALUE
Description This command sets the measurement level that indicates a
trigger stop condition. The power meter will stop taking further
measurements if a measurement taken exceeds the value specified by this
command, and if trigger stop option is set to measurement level.
Syntax PM:TRIG:VALUE value
Parameters
The parameter <value> is of type <number>
Argument Value Description
value float Measurement level
Related Commands: PM:TRIG:STOP
8.3.3.98 PM:TRIG:VALUE?
Description This command returns the measurement level that indicates
a trigger stop condition.
Syntax PM:TRIG:VALUE?
Parameters None

Table of Contents

Other manuals for Newport 1936-R series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Newport 1936-R series and is the answer not in the manual?

Newport 1936-R series Specifications

General IconGeneral
BrandNewport
Model1936-R series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals