Communication Command Reference 85 85
8.3.3.5 ADDRess?
Description USB address query.
Syntax ADDRess?
Remarks The ADDRess query returns the power meter’s USB
address.
Response Value Description
address 0 Reserved
1 to 31 Valid USB address range
Related Commands: ADDRess
8.3.3.6 BEEP
Description Beep command
Syntax BEEP beep set
Remarks The BEEP command controls the power meter’s beeper.
The beeper can be used to signal error or warning
conditions.
Response Value Description
beep set 0 Disable beeper
1 Enable beeper
2 Beep for 100 ms
Note: The beeper is enabled at power on.
Related Commands: BEEP?
8.3.3.7 BEEP?
Description Beep query
Syntax BEEP?
Remarks The BEEP? query returns the enable status of the power meter’s
beeper.
Response Value Description
beep set 0 Beeper disabled
1 Beeper enabled
Related Commands: BEEP
8.3.3.8 ECHO
Description Echo command
Syntax ECHO echo set