92 Communication Command Reference
8.3.3.28 PM:CORR?
Description Power measurement correction settings query
Syntax PM:CORR?
Remarks The PM:CORR? command returns the power
measurement correction settings. These settings are used by
the power meter to correct the actual power measurement.
The corrected power is calculated using the formula
provided below:
Corrected Measurement =
((Actual Measurement * value1) + value2) * value3
Response Type Description
value1 float Multiplier 1 (default = 1.000)
value2 float Offset (default = 0.000)
value3 float Multiplier 2 (default = 1.000)
Related Commands: PM:CORR, PM:Power?, PM:DPower?
8.3.3.29 PM:DETMODEL?
Description Detector model query
Syntax PM:DETMODEL?
Remarks The PM:DETMODEL? query returns the model number
of the detector.
For example: 818-SL.
Response Type Description
model string Detector model number
8.3.3.30 PM:DETSIZE?
Description Detector surface area
Syntax PM:DETSIZE?
Remarks The PM:DETSIZE? query returns the surface area of the
detector in cm
2
.
Response Type Description
Area float Detector surface area
8.3.3.31 PM:DETSN?
Description Detector serial number query
Syntax PM:DETSN?

Table of Contents

Other manuals for Newport 1936-R series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Newport 1936-R series and is the answer not in the manual?

Newport 1936-R series Specifications

General IconGeneral
BrandNewport
Model1936-R series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals