140 Appendix B
This error is generated if the command in the input buffer is too large to fit into
the available space in the parser buffer. It is usually generated when commands
are sent to the instrument faster than it can process. Parser buffer is 2,048
characters long.
12.4 Device Errors
Device Errors are associated with some system condition that affects the
operation of the meter. Errors associated with data reading will set the
appropriate bit but will not generate an error message to avoid jamming the
error queue or the interface.
701, Detector Calibration Read or Write Failed.
An error was encountered during a read/write operation to the calibration
EPROM in the detector. If problem persists, contact the factory.
703, Power Meter set to defaults due to Firmware update
After upgrading the firmware, in some cases the changes are significant
enough to require resetting the instrument to factory defaults.
704, User reference cannot be changed/stored while you are in units of
Watts or dBm. Change to dB or Rel to set the user reference value.
User reference can only be stored in dB or Rel mode.
705, Illegal data store parameter change. Queue cleared.
An action that affects the data store caused the data store queue to be cleared.
706, Digital Filter Interval changed, must not be greater than Data Store
interval.
The digital filter interval cannot be greater than the data store interval.
707, Digital Filter Disabled with External Trigger.
Digital filter must be disabled during external trigger mode.
708, There is no new data for a statistics update.
No new data has been stored in the data store since the last statistics update.
709, Statistics are not calculated while Data Store is running.
User attempted to compute statistics while data store was active. Turn off
data store and compute.