114 Communication Command Reference
Related Commands: PM:TRIG:STOP
8.3.3.99 PM:TRIG:TIME
Description This command sets the time duration that indicates a trigger
stop condition. The power meter will stop taking further measurements if the
time from trigger start exceeds time duration specified by this command, and
if trigger stop option is set to time.
Syntax PM:TRIG:TIME value
Parameters
The parameter <value> is of type <number>
Argument Value Description
value int Time (milli-seconds)
Related Commands: PM:TRIG:STOP
8.3.3.100 PM:TRIG:TIME?
Description This command returns the time duration that indicates a
trigger stop condition.
Syntax PM:TRIG:TIME?
Parameters None
Related Commands: PM:TRIG:STOP
8.3.3.101 PM:UNITs
Description Units select
Syntax PM:UNITS units
Remarks The PM:UNITS command selects the units for readings.
Argument Value Description
units 0 Amps
1 Volts
2 Watts
3 Watts/cm
2
4 Joules
5 Joules/cm
2
6 dBm
7-10 Reserved
11 Sun
Related Commands: PM:MODE?

Table of Contents

Other manuals for Newport 1936-R series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Newport 1936-R series and is the answer not in the manual?

Newport 1936-R series Specifications

General IconGeneral
BrandNewport
Model1936-R series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals