84 Communication Command Reference
bin 0 Reserved
1 to 5 Valid configuration settings
Related Commands: *SAV
8.3.3.3 *SAV
Description Save Configuration Settings
Syntax *SAV bin
Remarks The *SAV command saves the present state of the power
meter in its non-volatile flash memory. A particular state is
then recalled using the *RCL command. If any one of
these parameters are changed, the present state of the power
meter will automatically be saved in configuration setting
#1. When the power meter is reset, the state of the meter
defaults to configuration setting #1. The setup parameters
saved include:
Display brightness level
USB address
Color scheme
Measurement display mode
Custom wavelengths
Argument Value Description
bin 0 Reserved
1 to 5 Valid configuration settings
Related Commands: *RCL
8.3.3.4 ADDRess
Description USB address command.
Syntax ADDRess value
Remarks The ADDRess command sets the power meter USB
address. After changing USB address, the communication
with the power meter has to be re-initialized.
Argument Value Description
Value 0 Reserved
1 to 31 Valid USB address range
Related Commands: ADDRess?