Communication Command Reference 93 93
Remarks The PM:DETSN? query returns the serial number of the
detector.
For example: 0001
Response Type Description
serial number string Detector serial number
8.3.3.32 PM:DIGITALFILTER
Description Digital filter select command
Syntax PM:DIGITALFILTER value
Remarks The PM:DIGITALFILTER command specifies the digital
filter setting.
Argument Type Description
Value int digital filter size between 0 and 10000
Related Commands: PM:DIGITALFILTER?
8.3.3.33 PM:DIGITALFILTER?
Description Digital filter query
Syntax PM:DIGITALFILTER?
Remarks The PM:DIGITALFILTER? query returns the present
digital filter setting.
Argument Type Description
Value int digital filter size between 0 and 10000
Related Commands: PM:DIGITALFILTER
8.3.3.34 PM:DPower?
Description Detector power query
Syntax PM:DPower?
Remarks The PM:DPower? returns the actual power measurement.
This measurement does not include any correction settings
specified using “PM:CORR” command. When the
correction settings are set to default values, the power
measurement returned by this command is same as the
measurement returned by “PM:P?” command.
Response Type Description
Power float Power in Watts
Related Commands: PM:Power?

Table of Contents

Other manuals for Newport 1936-R series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Newport 1936-R series and is the answer not in the manual?

Newport 1936-R series Specifications

General IconGeneral
BrandNewport
Model1936-R series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals