EasyManua.ls Logo

Olivetti PR2 plus - CIS Assy (Upper Scanner) Disassembly;Reassembly

Olivetti PR2 plus
172 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
XZAC4219-05 Service Manual
7-33
CIS assy (upper scanner) disassembly/reassembly
1. Remove the casing. See the section “Casing disassembly/reassembly” on page 7 - 4.
2. Remove the upper mechanical assy. See the section “Upper mechanical assy disassembly/reassembly” on
page 7 - 11.
3. Remove the roller assy. See the section “Roller assy (upper scanner) disassembly/reassembly” on page 7 - 31.
4. Remove the two snap rings (1).
5. Remove the CIS assy (2) lifting it and taking care not to mislay the spring (3).
Figure 7 - 47 Removing the CIS unit
6. Disconnect the cable (4) from the CIS (2) taking great care not to damage the connectors (use a slotted
screwdriver as indicated in the figure).
Figure 7 - 48 Disconnecting the CIS cable
3
2
1
1
4
2

Table of Contents

Other manuals for Olivetti PR2 plus

Related product manuals