EasyManua.ls Logo

Olivetti PR2 plus - CIS Assy (Lower Scanner) Disassembly;Reassembly

Olivetti PR2 plus
172 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7-34 Service Manual
XZAC4219-05
NOTE: for reassembly, reverse the procedure, paying particular
attention to correct fitting of the spring (3).
See the Figure 7 - 47.
Clean the glass of the CIS with a soft cloth.
After completing the re-assembly procedure, carry out
mechanical adjustment and calibration of the scanner
(see the section “Calibrating the scanner” on page 4 - 28).
CIS assy (lower scanner) disassembly/reassembly
1. Disassembling the casing. See the section “Casing disassembly/reassembly” on page 7 - 4.
2. Remove the upper mechanical assy. See the section “Upper mechanical assy disassembly/reassembly” on
page 7 - 11.
3. Remove the roller guard (1) removing the related M3 lock-screws (2) and release the cable (3) from the cable
clips (4).
Figure 7 - 49 Disassembling the CIS assy (lower scanner)
3
2
1
4

Table of Contents

Other manuals for Olivetti PR2 plus

Related product manuals