Appendix A: Troubleshooting
Page A-6 Lasair II Particle Counter Operations Manual
7 If using Internet Explorer, and web page won't load properly, press CTRL key
while selecting “REFRESH” to force page reload over the network.
P. Data Cannot be Erased
8 Data cannot be erased while statistics mode is on.
9 Data cannot be erased from the Web Server in the following conditions:
• if the instrument is being run from front panel
• if LCD screen is on one of the setup screens.
Q. Facility Net will not connect to the particle counter.
10 Ensure that “Enable Fac-Net” is checked on the Comm Setup Screen.
11 Ensure that the Lasair II particle counter display is set to Main Screen.