Serial Number Output to Pharmaceutical Net
Lasair II Particle Counter Operations Manual Page H-3
4 Press the Return Arrow icon softkey to save the check box status and exit
the Edit mode. All applicable output will include 0.5μm and 5.0μm data
only.
NOTE:
If the TCP/IP setting is not selected, the LCD will display two
channels but it will also report all six channels to Pharmaceutical Net.
Serial Number Output to Pharmaceutical Net
The Lasair II particle counter serial number is sent to Pharmaceutical Net during
initial communications over TCP/IP. It is not, however, stored.
Print Output
This special pharmaceutical mode output applies to standard printouts and statistics
average data printouts.
It does not apply to the following statistical outputs:
• FS-209
• ISO
• BS-5295
It also does not apply to the following other outputs:
• TCP/IP protocol data transfer (unless enabled as described in “Enabling the
Pharmaceutical Two-Channel Output Mode” on page H-1.)
• CSV data output
• HTML data output
The following illustrates the format and content of the two-channel (0.5 and
5.0 μm) print output:
Final Sample Normalized
LASAIR II
006 Invalid
2001/04/13 12:40:48#1
|| μ || Δ || ∑ ||
|| 0.5|| 713777|| 1011512||
|| 5.0|| 30024|| 46108||