Lasair II Particle Counter Operations Manual Page 1-1
Chapter 1
Getting Started
Product Overview
The Lasair II particle counter is an airborne-particle counter designed for either
mobile or dedicated use. The sizing channel threshold and the sample flow rate vary
with the particular model. See the chart below for each model’s channel thresholds
and flow rates.
Counting Technique
When the Lasair II particle counter begins sampling, the internal pump creates a
vacuum to pull air into the sample inlet. The sample flow is controlled to maintain
1 CFM (50 LPM for the 350L and 550L) of airflow through the sampler. A variance
of only +5% in airflow is allowed in a valid sample.
Model
No.
Channel Thresholds Flow Rate Availabl
e
350L 0.3, 0.5, 1.0, 5.0, 10.0, 25.0 µm 1.77 cfm (50 LPM) Yes
550L 0.5, 1.0, 2.0, 5.0, 10.0, 25.0 µm 1.77 cfm (50 LPM) Yes
310A 0.3, 0.5, 1.0, 5.0, 10.0, 25.0 µm 1.0 cfm (28.3 LPM) Yes
310B 0.3, 0.5, 1.0, 3.0, 5.0, 10.0 µm 1.0 cfm (28.3 LPM) Yes
510A 0.5, 1.0, 2.0, 5.0, 10.0, 25.0 µm 1.0 cfm (28.3 LPM) Yes
310 0.3, 0.5, 0.7, 1.0, 5.0, 10.0 µm 1.0 cfm (28.3 LPM) No
325 0.3, 0.5, 1.0, 5.0, 10.0, 25.0 µm 1.0 cfm (28.3 LPM) No
510 0.5, 0.7, 1.0, 2.0, 5.0, 10.0 µm 1.0 cfm (28.3 LPM) No
525 0.5, 0.7, 1.0, 5.0, 10.0, 25.0 µm 1.0 cfm (28.3 LPM) No

Table of Contents

Other manuals for Particle Measuring Systems LASAIR II 310

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Particle Measuring Systems LASAIR II 310 and is the answer not in the manual?

Particle Measuring Systems LASAIR II 310 Specifications

General IconGeneral
BrandParticle Measuring Systems
ModelLASAIR II 310
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals