Multi-Drop Command Coding Overview
Lasair II Particle Counter Operations Manual Page F-3
Multi-Drop Commands
CCAL
Response: "RCAL yyyy/mm/dd"
Returns the last calibration date. (A Lasair II particle counter
should be calibrated at least once a year).
CCONT n
Response: “RCONT m
CDELAY sec
Response: “RDELAY sec
This is a backward compatible command. Delay is specified as start
of first sample to start of next sample. Sample volume or sample
time must be set first for this command to work properly. Times are
approximate if sample volume is set rather than sample time.
Set continuous sampling on/off (1/0). If turned off, then number of
samples is set to one. Use CNOS to set number of samples.
CDT yy mm dd hh mm ss
Set date and time. Either two digit or four digit year is ok.
Response: “RDT
CES
End Sampling.
Response: “RES 1 (success) or RES 0” (failure)
CFQ
Response: “RPQ 1 or RPQ 0
Flush data queue--erase all data. Note: The particle counter can un-
erase data from this command if no samples are taken, and
locations are not erased.
CID name
Response: “RID name
Set sample ID (location) to given string. This command translates
all characters to upper case. The only legal characters in this
command are
A-Z, a-z, and 0-9. Other characters which are
legal for the Lasair II particle counter location names are not legal
here.
CISDELAY sec
Response: “RISDELAY sec
Set inter-sample delay in seconds. Delay is from end-to-start.
CKB n
Response: “RKB m

Table of Contents

Other manuals for Particle Measuring Systems LASAIR II 310

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Particle Measuring Systems LASAIR II 310 and is the answer not in the manual?

Particle Measuring Systems LASAIR II 310 Specifications

General IconGeneral
BrandParticle Measuring Systems
ModelLASAIR II 310
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals