Appendix E: Comma Separated Value (CSV) Format
Page E-2 Lasair II Particle Counter Operations Manual
13
Sample Valid: The sample will be declared invalid if the laser current is bad
during the sample, or if the sample flow falls outside the 5% ISO limit, or if the
user invalidates the sample manually (only under statistics mode).
14
Sample Interval: In time. Format is HH:MM:SS
15
Volume Units: CF or CM
16
Sample Volume: Floating point value. In "Volume Units" given above.
17
Number Channels: Currently 6.
18
Size 1: This is fixed for the instrument
19
Counts 1: The total differential raw count value.
20
Size 2: This is fixed for the instrument
21
Counts 2: ditto
22
Size 3: This is fixed for the instrument
23
Counts 3: ditto
24
Size 4: This is fixed for the instrument
25
Counts 4: ditto
26
Size 5: This is fixed for the instrument
27
Counts 5: ditto
28
Size 6: This is fixed for the instrument
29
Counts 6: ditto
30
Number Analog: This can vary from 0 to 6 depending upon what is enabled on
the instrument. To prevent column alignment issues when adding analog channels,
all unused physical channels that are before the last used channel will be reported.
Thus, the value reported here includes those unused channels which are before the
last used channel.
31
Name 1: User name (16 char max) or dedicated temperature probe. Name of temp.
probe is currently "T". If the first character of the name is a "?", that indicates the
name has changed since the data sample was saved. (The data sample only has an
index to the name.)
32
Units 1: User name (10 char max). Might be dedicated to temperature sensor if
probe enabled. If the first character of the units is a "?", that indicates the units has
changed since the data sample was saved. (The data sample only has an index to
the units.)
CSV Value Format and Description (Continued)
Ref.
No
Data Description

Table of Contents

Other manuals for Particle Measuring Systems LASAIR II 310

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Particle Measuring Systems LASAIR II 310 and is the answer not in the manual?

Particle Measuring Systems LASAIR II 310 Specifications

General IconGeneral
BrandParticle Measuring Systems
ModelLASAIR II 310
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals