Specifications
Lasair II Particle Counter Operations Manual Page 1-9
Supporting Software Facility Net, Pharmaceutical Net, Internet Explorer
®
External Surface ESD-compatible KYDEX
®
Tested
Cleaning Materials
Bleach, formaldehyde, ethyl or isopropyl alcohol, peroxide/
quaternary ammonium solutions, DECON-Clean
®
, DECON-
QUAT 100
®
, HYPO-CHLOR
®
, STERI-PEROX
®
, and
PHENE-AHOL
®
spray disinfectant
Sample Output Internally filtered, HEPA standard (> 99.97% at 0.3 μm)
Display 1/4 VGA color LCD screen
Languages English, Chinese (Traditional and Simplified), French,
German, Italian, Japanese (Kanji), Korean, Spanish
Printer Built-in thermal printer
Reports Sample printouts; final reports for EU GMP,
ISO 14644-1, FS-209E, and BS-5295
Power 100–240 V ±10%, 50–60 Hz, 1.25 A maximum
Battery (optional) Removable NiMH battery, 16.8 V nominal voltage
Dimensions (H,W,D) 11.5 × 10.7 × 9.0 inches (29 × 27 × 23 cm)
Weight 12.0 lbs. (5.4 kg.) without battery; 15.5 lbs. (7 kg.) with
optional battery
Operating
Environment
Temperature 15–35°C; humidity: 10–90%, non-condensing
Storage or
transportation
environment
Temperature: − 40 to 70°C
Humidity: 5–90% non-condensing
Indoor/outdoor Indoor use only-not protected against harmful ingress of
moisture
Class Class I Equipment (requires a grounded type electrical supply)
Installation
over-voltage
Category II for transient over-voltage
Product laser
classification
Class I, when used as designed