Prma 460 126 User Manual
Appendix 8 Electromagnetc compatblty (EM)
Electromagnetic environment
The Prma 460 anaesthetc machne s sutable for use n the specfed electromagnetc envronment The user of the Prma
460 anaesthetc machne should assure that t s used n an electromagnetc envronment as descrbed below
hanges or modfcatons to ths anaesthetc machne system, not expressly approved by Penlon Lmted, could result n EM
ssues wth ths anaesthetc machne ontact Penlon Lmted for more nformaton
The use of phones or RF emttng equpment near ths anaesthetc machne may cause nterference
Always montor anaesthetc machne operaton before and durng use on a patent
The essental performance of the anaesthetc machne s to provde controlled concentratons and flows of anaesthesa gases
nto a patent breathng system
NOTE
Masmo NomoLne ISA AX+ and ISA OR+ modules - refer to Appendx 7
WARNIN
The devce should not be used adacent to or stacked wth other manufacturer’s equpment If adacent or stacked use s
necessary, the devce should be observed to verfy normal operaton n the confguraton n whch t wll be used
ables, Transducers, and Accessores
WARNIN
The devce s EM-complant wth all cables, transducers and accessores suppled by Penlon Lmted
The use of cables, transducers and accessores other than those specfed may result n ncreased emssons or decreased
mmunty of ths devce
The use of cables, transducers and accessores suppled by Penlon Lmted on non-Penlon equpment may also result n
ncreased emssons or decreased mmunty of that equpment
ables suppled by Penlon Lmted
Category
Function Length (m)
O sensor cable (fuel cell sensor type) Connects sensor to machine 0.25
Absorber cable Connects spirometers to machine 0.8
udance and manufacturer’s declaraton – electromagnetc emssons
The Prma 460 anaesthetc machne s ntended for use n the electromagnetc envronment specfed below
The customer or user should assure that t s used n such an envronment
Required test Compliance omments
RF emssons
ISPR 11
roup 1
lass A
lass A equpment s sutable for use n all locatons other than those allocated n
resdental envronments and those drectly connected to a low voltage power supply
network whch supples buldngs used for domestc purposes, and may be used n
domestc establshments and those drectly connected to the publc low-voltage power
supply network that supples buldngs used for domestc purposes provded the followng
warnng s heeded
NOTE The EMISSIONS characterstcs of ths equpment make t sutable for use n
ndustral areas and hosptals (ISPR 11 class A) If t s used n a resdental envronment
(for whch ISPR 11 class B s normally requred) ths equpment mght not offer adequate
protecton to rado-frequency communcaton servces The user mght need to take
mtgaton measures, such as relocatng or re-orentng the equpment
Harmonc emssons
IE 61000-3-2
omples
Voltage fluctuatons/
flcker emssons
IE 61000-3-3
omples
Appendix 8