Description
3.14.5 Oxygen Monitor Alarms
HIH O ALARM
The hgh O alarm s trggered when the oxygen concentraton
s 1% above the set value
The hgh O Alarm vsual ndcator wll llumnate
A hgh prorty audble alarm wll sound
To cancel ths alarm, the hgh alarm settng must be equal to,
or above the oxygen concentraton
The alarm can be muted for 120 seconds
LOW O ALARM
The low alarm s trggered when the oxygen concentraton s
1% below the set value
The Low O Alarm vsual ndcator wll llumnate
A hgh prorty audble alarm wll sound
To cancel ths alarm, the low alarm settng must be equal to,
or below the oxygen concentraton
The alarm can be muted for 120 seconds
O SENSOR FAULT
The alarm s trggered
1 When ether the oxygen sensor s dsconnected or
approachng the end of ts lfe
2 If the O concentraton exceeds 110%
The message O SENSOR FAULT wll be dsplayed
A hgh prorty audble alarm wll sound
To cancel ths alarm, check the sensor connecton or replace
the sensor
The alarm can be muted for 120 seconds
O SENSOR LOW
The sensor has approached the end of ts lfe
The legend O SENSOR LOW wll be dsplayed, and a low
prorty alarm (sngle note) wll sound
The sensor must be replaced as the output wll fall very quckly
to zero wthn two to three weeks of normal usage
See secton 831 for sensor replacement
3.14.6 Oxygen Monitor Alarm Mute
In an alarm condton, pressng the ALARM MUTE button (1)
wll deactvate the audble alarm, but the alarm message
dsplay wll reman on screen
The swtch wll llumnate, and a sngle note wll sound
The alarm mute cannot be operated
1 Untl the mute tme s over, or the alarm condton has
been rectfed
2 When oxygen concentraton drops below 18%
1