Prma 460 32 User Manual
Description
3.12.7 Configuration
31271 Start-up Screens
1 Start-up
a) If requred, Touchscreen calbraton can be ntated
from the ntroducton screen
b) The ntroducton screen allows the user to select one
of three default settngs
• SITE DEFAULT
• ADULT DEFAULT
• PAEDIATRI DEFAULT
NOTE
1 The user must select one of the above default groups
before the ventlator wll swtch to standby n that default
mode
2 SITE DEFAULT can be edted n standby mode
3 Settngs can be saved va the servce menu to create a new
ste default
Functon Ste Default Adult Default Paedatrc Default
VT (mL) 20 - 1600 600 150
VM (L/mn) 05 - 50 60 22
T+PS INIT (cmHO) 4 - 70 10 10
Set BPM 4 - 100 10 15
I  E 102 to 18 12 12
PEEP (cmHO) Off 4 - 20 (opton Off 4 - 30) OFF OFF
Lmt (cmHO) 38 38 38
Trgger (L/mn) 1 1 1
Apnoea alarm mute lmt (s)
Adustable n Spontaneous
mode only
15, 30, 60, 120, 180 15 15
Volume type Tdal or Mnute Tdal Tdal
Back lght level (%) 0 - 100 100 100
3.12.7.2 Default Settings Selection
The user can select ADULT, or PAEDIATRI, or SITE, and vew
the default parameter settngs The optons wll reman, even
after the ventlator s turned off
3.12.7.3 Site Default Settings
The values lsted are examples
Adust the parameter values from wthn the Ste Defaults
menu (1)
Press to confrm the new settngs for ste defaults
1