Prma 460 32 User Manual
Description
3.12.7 Configuration
31271 Start-up Screens
1 Start-up
a) If requred, Touchscreen calbraton can be ntated
from the ntroducton screen
b) The ntroducton screen allows the user to select one
of three default settngs
• SITE DEFAULT
• ADULT DEFAULT
• PAEDIATRI DEFAULT
NOTE
1 The user must select one of the above default groups
before the ventlator wll swtch to standby n that default
mode
2 SITE DEFAULT can be edted n standby mode
3 Settngs can be saved va the servce menu to create a new
ste default
Functon Ste Default Adult Default Paedatrc Default
VT (mL) 20 - 1600 600 150
VM (L/mn) 05 - 50 60 22
T+PS INIT (cmHO) 4 - 70 10 10
Set BPM 4 - 100 10 15
I E 102 to 18 12 12
PEEP (cmHO) Off 4 - 20 (opton Off 4 - 30) OFF OFF
Lmt (cmHO) 38 38 38
Trgger (L/mn) 1 1 1
Apnoea alarm mute lmt (s)
Adustable n Spontaneous
mode only
15, 30, 60, 120, 180 15 15
Volume type Tdal or Mnute Tdal Tdal
Back lght level (%) 0 - 100 100 100
3.12.7.2 Default Settings Selection
The user can select ADULT, or PAEDIATRI, or SITE, and vew
the default parameter settngs The optons wll reman, even
after the ventlator s turned off
3.12.7.3 Site Default Settings
The values lsted are examples
Adust the parameter values from wthn the Ste Defaults
menu (1)
Press to confrm the new settngs for ste defaults
1