Prma 460 87 User Manual
Operating the Anaesthesia System
7.3.4 Weekly Checklist
At least every week, n addton to the daly functon test, the
followng checks must be carred out
Alarms
1 Select STANDBY MODE
2 Unplug the mans power cable from the A outlet
The MAINS FAILURE alarm should actvate
3 Reconnect the mans power cable to the A outlet The
alarm should turn off
NOTE
Do not use the system f there s any malfuncton
If the problem cannot be rectfed, the system must be checked
by a traned engneer
Bellows
heck the condton of the bellows (1) and exhalaton
daphragm valve (2)